User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HEA 4934A/4935A TRANSMITTER FLATNESS CALIBRATION DATE: 20-Jul-93 AUTHOR: User Contributed REVISION: 1 F. Samonski 18-May-93 Added 4934A ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 87 CONFIGURATION: Fluke 8506A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R D N B P J S M F W 1.002 ASK+ X 1.003 HEAD [27][91]1m {TRANSMITTER FLATNESS CALIBRATION +10dBM} 1.004 DISP 1.004 DISP Connect the UUT TRMT/RCV phone jack to the 8506A V/OHM 1.004 DISP INPUT terminals. Terminate the 8506A HI and LOW INPUTS 1.004 DISP with a precision 600 ohm load. 1.004 DISP 1.005 DISP Set UUT controls as follows: 1.005 DISP DISPLAY TRMT 1.005 DISP MEASUREMENT LEVEL FREQUENCY 1.005 DISP SETUP TRMT 600 1.005 DISP REV OFF (NOR 4935A) 1.005 DISP BRG, DIAL, & HOLD OFF 1.005 DISP IMPULSE NOISE DISPLAY (1004Hz) 1.006 DISP 1.006 DISP Adjust the UUT OUTPUT LEVEL for a +10 dBm on UUT. 1.006 DISP 1.007 8506 10 2.435V 0.005U 1004H E N 2W 1.008 DISP 1.008 DISP Maintain a +10 dBm indication on the UUT for the 1.008 DISP remainder of the test. 1.009 DISP 1.009 DISP Set the UUT FREQUENCY controls for a 20Hz display. 1.010 8506 10 V 0.295U 20H E C 2W 2.001 DISP 2.001 DISP Set the UUT FREQUENCY controls for a 200Hz display. 2.002 8506 10 V 0.055U 200H E C 2W 3.001 DISP 3.001 DISP Set the UUT FREQUENCY controls for a 5KHz display. 3.002 8506 10 V 0.055U 5kH E C 2W 4.001 DISP 4.001 DISP Set the UUT FREQUENCY controls for a 15KHz display. 4.002 8506 10 V 0.055U 15kH E C 2W 5.001 DISP 5.001 DISP Set the UUT FREQUENCY controls for a 30KHz display. 5.002 8506 10 V 0.055U 30kH E C 2W 6.001 DISP 6.001 DISP Set the UUT FREQUENCY controls for a 60KHz display. 6.002 8506 10 V 0.145U 60kH E C 2W 7.001 DISP 7.001 DISP Set the UUT FREQUENCY controls for a 85KHz display. 7.002 8506 10 V 0.145U 85kH E C 2W 8.001 HEAD 8.002 HEAD [27][91]1m {TRANSMITTER FLATNESS CALIBRATION -40dBM} 8.003 DISP 8.003 DISP Set the TI Frequency to 1004hz 8.003 DISP (Press Impulse Noise Display button once) 8.003 DISP 8.004 DISP Adjust the UUT OUTPUT LEVEL for a -40 dBm on UUT. 8.004 DISP 8.005 8506 100 7.746mV 1U 1004H E N 2W 8.006 DISP 8.006 DISP Maintain a -40 dBm indication on the UUT for the 8.006 DISP remainder of this test. 8.007 DISP 8.007 DISP Set the UUT FREQUENCY controls for a 20Hz display. 8.008 8506 100 mV 0.934U 20H E C 2W 9.001 DISP 9.001 DISP Set the UUT FREQUENCY controls for a 50Hz display. 9.002 8506 100 mV 0.934U 50H E C 2W 10.001 DISP 10.001 DISP Set the UUT FREQUENCY controls for a 200Hz display. 10.002 8506 100 mV 0.174U 200H E C 2W 11.001 DISP 11.001 DISP Set the UUT FREQUENCY controls for a 15KHz display. 11.002 8506 100 mV 0.174U 15kH E C 2W 12.001 DISP 12.001 DISP Set the UUT FREQUENCY controls for a 60KHz display. 12.002 8506 100 mV 0.454U 60kH E C 2W 13.001 DISP 13.001 DISP Set the UUT FREQUENCY controls for a 85KHz display. 13.002 8506 100 mV 0.454U 85kH E C 2W 14.001 END